Universal R-320 Operations Instructions Page 104

  • Download
  • Add to my manuals
  • Print
  • Page
    / 117
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 103
DE
EN
104/117
1619
1462-A
--- SK 56.04-4
0509 0513
Ausschwingrotor 6-fach /
Swing out rotor 6-times
90°
Kapazität / capacity ml
15 50 12 8 8,5 – 10 7,5-8,2 9-10 10
Maße / dimensions x L
mm
17 x 120 29 x 115 16 x 101 16 x 125 16 x 100 15 x 92 16 x 92 15 x 102
Anzahl p. Rotor / number p. rotor
6 6 6 6 6 6 6 6
Drehzahl / speed RPM
4000 4000 4000 4000 4000 4000 4000 4000
RZB / RCF
2701 2701 2647 2647 2647 2647 2647 2647
Radius / radius mm
151 151 148 148 148 148 148 148
9 (97%)
sec
20
9
sec
22
Temperatur / temperature °C
1)
- 15
Probenerwärmung/Sample
temp. rise
K
2)
+ 9
1617
1462-A
---
0509 0513
Ausschwingrotor 8-fach /
Swing out rotor 8-times
45°
Kapazität / capacity ml
15 50
Maße / dimensions x L
mm
17 x 120 29 x 115
Anzahl p. Rotor / number p. rotor
8 8
Drehzahl / speed RPM
5000 5000
RZB / RCF
3857 3857
Radius / radius mm
138
9 (97%)
sec
20
9
sec
19
Temperatur / temperature °C
1)
- 10
Probenerwärmung/Sample
temp. rise
K
2)
+ 14
1) Tiefste erreichbare Temperatur bei maximaler Drehzahl, 1 h
Laufzeit und 20°C Raumtemperatur (nur bei Kühlzentrifuge)
1) Lowest possible temperature during maximum speed, 1 h running time and
20°C ambient temperature (only with cooling centrifuges)
2) Probenerwärmung bei maximaler Drehzahl und 1 Stunde
Laufzeit (nur bei Zentrifuge ohne Kühlung)
2) Sample temp. rise during maximum speed and 1 hour running time
(only with centrifuges without cooling)