Universal R-320 Operations Instructions Page 45

  • Download
  • Add to my manuals
  • Print
  • Page
    / 117
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 44
EN
45/117
26 Faults
If the fault cannot be eliminated with the help of the fault table, please inform Customer Service.
Please specify the type of centrifuge and the serial number. Both numbers can be found on the name plate of the
centrifuge.
Perform a MAINS RESET:
Switch off the mains switch (switch position "0").
Wait at least 10 seconds and then switch on the mains switch again (switch position "Ι").
Message / fault Cause Remedy
No display --- No voltage.
Overvoltage protection tripped out.
Check supply voltage.
Mains switch ON.
TACHO - ERROR 1, 2, 96 Faulty speedometer.
Motor, electronics defective.
CONTROL - ERROR 8 Error in lid locking or lid closure.
Open the cover.
Switch off the mains switch (switch
position "0").
Wait at least 10 seconds.
Turn the rotor vigorously by hand.
Switch on the mains switch again
(switch position "Ι"). The rotor must
turn during switch-on.
IMBALANCE --- The rotor is unevenly loaded.
Open lid.
Check the loading of the rotor, see
chapter "Loading the rotor".
Repeat the centrifugation run.
CONTROL - ERROR 4, 6 Error in lid locking or lid closure.
N > MAX 5 Rotation too fast
N < MIN 13 Rotation too slow
Perform a MAINS RESET.
MAINS INTERRUPT --- Power failure during the centrifugation run.
(The centrifugation run was not finished.)
Open lid.
Push
START / IMPUL
S
button.
Repeat the centrifugation run if
necessary.
ROTORCODE 10.1,
10.2
Incorrect rotor coding
Open lid.
CONTROL-ERROR 21, 22,
25, 27,
29
Error / defect electronics
CONTROL-ERROR 23 Error / defect controls
SER I/O - ERROR 30, 31,
33, 36
Error / defect electronics
° C * - ERROR 51 - 53,
55
Error / defect electronics
FU / CCI - ERROR 60 - 64,
67, 68,
82 - 85
Error / defect electronics / motor
SYNC-ERROR 90 Error / defect electronics
SENSOR-ERROR 91 - 93 Error / defect unbalance sensor
KEYBOARD-ERROR --- Error / defect controls
Perform a MAINS RESET.
NO ROTOR --- No rotor installed
Open lid.
Install rotor.
N > ROTOR MAX
--- Speed in the selected program greater than
the maximum speed of the rotor.
Check the set speed.
Reduce the set speed
Page view 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 116 117

Comments to this Manuals