Universal R-320 Operations Instructions Page 29

  • Download
  • Add to my manuals
  • Print
  • Page
    / 117
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 28
EN
29/117
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model UNIVERSAL 320 R
Type 1406-50 1406-51 1406-70 1406-71
Mains voltage (± 10%)
200-240 V
1 100-127 V 1 200-240 V 1 100-127 V 1
Mains frequency 50 – 60 Hz 50 – 60 Hz 50 – 60 Hz 50 – 60 Hz
Connected load 400 VA 400 VA 400 VA 400 VA
Current consumption 2.0 A 4.0 A 2.0 A 4.0 A
Max. capacity 4 x 100 ml
Allowed density
1.2 kg/dm
3
Speed (RPM) 15000
Force (RCF) 21382
Kinetic energy 8600 Nm
Obligatory inspection (BGR 500) no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 35°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
ΙΙ
Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
EN / IEC 61326-1,
Class B
FCC Class B
EN / IEC 61326-1,
Class B
FCC Class B
Noise level (dependent on rotor)
58 dB(A)
Dimensions
Width
395 mm
Depth
520 mm
Height
346 mm
Weight approx. 35 kg