EN
30/117
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model UNIVERSAL 320 R
Type 1406-20 1406-21
Mains voltage (± 10%)
200-240 V 1∼ 240 V 1∼ 115-127 V 1∼
Mains frequency 50 Hz 60 Hz 60 Hz
Connected load 800 VA 950VA
Current consumption 4.0 A 8.0 A
Cooling medium R 404A
Max. capacity 4 x 100 ml
Allowed density
1.2 kg/dm
3
Speed (RPM) 15000
Force (RCF) 21382
Kinetic energy 8600 Nm
Obligatory inspection (BGR 500) no
Ambient conditions (EN / IEC 61010-1)
− Set-up site
Indoors only
− Altitude
Up to 2000 m above sea level
− Ambient temperature
5°C to 35°C
− Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
− Excess-voltage category
(IEC 60364-4-443)
ΙΙ
− Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
− Emitted interference,
Interference immunity
EN / IEC 61326-1, Class B FCC Class B
Noise level (dependent on rotor)
≤ 58 dB(A)
Dimensions
− Width
401 mm
− Depth
695 mm
− Height
346 mm
Weight approx. 52 kg