1 Use according to specification
The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG.
The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also
includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this
purpose. A different use or application over and above this is deemed not in accordance with the specifications. The
company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
2 Residual risks
The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3 Technical specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model UNIVERSAL 320 UNIVERSAL 320 R
Type 1401 1401-01 1406 1406-01
Mains voltage (± 10%)
200-240 V 1 100-127 V 1 200-240 V 1 240 V 1
115-127 V 1
Mains frequency 50 – 60 Hz 50 – 60 Hz 50 Hz 60 Hz 60 Hz
Connected load 400 VA 400 VA 800 VA 950 VA
Current consumption 2.0 A 4.0 A 4.0 A 8.0 A
Cooling medium ---- R 404A
Max. capacity 4 x 100 ml
Allowed density
1.2 kg/dm
3
Speed (RPM) 15000
Force (RCF) 21382
Kinetic energy 8600 Nm
Obligatory inspection (BGR 500) no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 35°C 5°C to 35°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
ΙΙ
Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
EN / IEC
61326-1,
Class B
FCC Class B
EN / IEC 61326-1, Class B
FCC Class B
Noise level (dependent on rotor)
68 dB(A) 58 dB(A)
Dimensions
Width
395 mm 401 mm
Depth
520 mm 695 mm
Height
346 mm 346 mm
Weight approx. 31 kg approx. 53 kg
approx. 52 kg